Mayo Clinic Interactive Breath-Hold Control System (IBC) FAQs

How do I turn the system on?
Be sure to review the appropriate sections of the operating manual. Each remote display and base unit has an on/off switch – all must be turned on in order for the system to operate.

The LED display is not changing with respiration – what should I check?
  1. If the LED display is not changing – first check to assure that the reference button has been activated. Press and release the reference button briefly at the base unit. Expand and contract the bellows to verify the display changes with the bellows movement. If successful, repeat this test with the blue reference button on each remote display.
  2. If the display is still not changing, it is possible that your bellows is not plugged in properly or is defective. Check the bellows connections while expanding and contracting the bellows after establishing a reference with the bellows expanded. Listen carefully for any noises that would indicate a leak in the system. Note that a second bellows was sent with each system, when in doubt replace the bellows with the second bellows and retry test A.
  3. If the system is fully charged and tests A or B cannot be accomplished successfully – contact Mayo Clinic Medical Devices for more information.

One remote display is not charged, can I still use the system?
Yes, the system can be used with just one or two remote displays. However, do not charge the remote displays while the base unit is in use for clinical care.

Will the unit function correctly when not on battery power?
While the unit will function accurately while being charged, it is not recommended that the system be used for clinical care unless the batteries are completely charged.

Can I purchase replacement parts?
Yes. Contact Mayo Clinic Medical Devices.